iio: pressure: mprls0025pa improve driver error resilience
Improve driver error resilience by ignoring the measurement if any of
the 3 error flags gets set while interacting with the sensor.
Based on the datasheet, in table 14 on page 14, the status byte
contains:
bit 5 busy flag - 1 if device is busy
bit 2 memory integrity/error flag - 1 if integrity test failed
bit 0 math saturation - 1 if internal math saturation has occurred
Co-developed-by: Andreas Klinger <ak@it-klinger.de>
Signed-off-by: Andreas Klinger <ak@it-klinger.de>
Signed-off-by: Petre Rodan <petre.rodan@subdimension.ro>
Link: https://lore.kernel.org/r/20231229092445.30180-6-petre.rodan@subdimension.ro
Signed-off-by: Jonathan Cameron <Jonathan.Cameron@huawei.com>