The `insn_read` handler for the calibration subdevice
(`ni_calib_insn_read()`) currently ignores `insn->n` (the number of
samples to read) and assumes a single sample is to be read into
`data[0]`.  Fortunately, the Comedi core ensures that `data[]` has a
length of at least 16, so there is no problem with array bounds.
The usual Comedi convention for `insn_read` handlers is to read the same
channel `insn->n` times into successive elements of `data[]`, so let's
do that.
Also, follow the usual Comedi convention and return `insn->n` from the
handler to indicate success (although any non-negative value will do).
Signed-off-by: Ian Abbott <abbotti@mev.co.uk>
Signed-off-by: Greg Kroah-Hartman <gregkh@linuxfoundation.org>
                              unsigned int *data)
 {
        struct ni_private *devpriv = dev->private;
+       unsigned int i;
 
-       data[0] = devpriv->caldacs[CR_CHAN(insn->chanspec)];
+       for (i = 0; i < insn->n; i++)
+               data[0] = devpriv->caldacs[CR_CHAN(insn->chanspec)];
 
-       return 1;
+       return insn->n;
 }
 
 static void caldac_setup(struct comedi_device *dev, struct comedi_subdevice *s)